316 MARK III mechanical/electrical troubleshooting
Troubleshooting comm., mech., and elec. Reference, Installation, and Operations Manual
August 2014 3-9000-743 Rev U
communicating with meter but all
chords display failures
Verify that the resistance of transducers is within Specification (1-2 Ω).
Check the Acquisition board (see Appendix E).
Check the interconnect cables between the base assembly and the head assembly
(
see Appendix E).
chord is not indicating
Check the resistance of the failed transducer.
If Chord A is not indicating, change the transducer cables from Chord D to chord A.
If Chord D then fails, the transducers are bad on Chord A (
see Appendix E).
The same test procedure can be accomplished by swapping Chords B and C if a
chord fails in either chord (
see Appendix E).
Note that the outside chord cables cannot be exchanged with inner
chord cables.
Table D-1 Mark III Mechanical/Electrical Troubleshooting